Evaluation of Renal Allograft Vasculature Using Non-contrast 3D Inversion Recovery Balanced Steady-state Free Precession MRA and 2D Quiescent-interval Slice-selective MRA

  • Ali Serhal1,* ,
  • Pascale Aouad1,
  • Muhamad Serhal2,
  • Ashitha Pathrose1,
  • Pamela Lombardi1,
  • James Carr1,
  • Ryan Avery1 and
  • Robert R. Edelman1,3
Exploratory Research and Hypothesis in Medicine   2021;6(3):90-98

doi: 10.14218/ERHM.2021.00011

Received:

Revised:

Accepted:

Published online:

 Author information

Citation: Serhal A, Aouad P, Serhal M, Pathrose A, Lombardi P, Carr J, et al. Evaluation of Renal Allograft Vasculature Using Non-contrast 3D Inversion Recovery Balanced Steady-state Free Precession MRA and 2D Quiescent-interval Slice-selective MRA. Explor Res Hypothesis Med. 2021;6(3):90-98. doi: 10.14218/ERHM.2021.00011.

Abstract

Background and objectives

Renal transplant patients often require periodic imaging to evaluate the transplant vessel anastomosis for potential vascular complications. The use of non-contrast enhanced magnetic resonance angiography (NCE-MRA) techniques is encouraged in these patients because they are at increased risk of nephrogenic systemic fibrosis (NSF) due to their renal insufficiency. This study aimed to evaluate the performance of two NCE-MRA techniques (three-dimensional [3D] balanced steady-state free precession [bSSFP] with inversion recovery and quiescent-interval slice-selective [QISS]) for the evaluation of renal allograft vasculature in patients with clinical suspicion, or Doppler ultrasound, or both of arterial anastomotic stenosis.

Methods

A total of 43 patients were included in this retrospective study. Two radiologists independently scored the images from 3D bSSFP and QISS MRA sequences for image quality and confidence in anastomosis interpretation, and the degree of stenosis at the arterial anastomosis. Correlations with digital subtraction angiography (DSA) were carried out when available. In addition, inter-rater agreement was calculated.

Results

In total, 43 patients underwent QISS and 3D bSSFP MRA. For QISS, all cases were adequate for evaluation. For 3D SSFP, 86% of cases were adequate for evaluation. There was a good-to-excellent inter-rater agreement for all scores and an excellent correlation between NCE-MRA and DSA results when available (12 patients).

Conclusions

QISS and 3D SSFP showed good inter-rater agreement for image quality and stenosis grade, with more cases being of adequate image quality that used QISS. Further study is required; however, NCE-MRA shows potential as a risk-free alternative to CTA and contrast-enhanced MRA (CE-MRA) for the evaluation of arterial anastomoses in renal transplant patients.

Keywords

Renal transplant artery stenosis, Non-contrast enhanced magnetic resonance angiography, Quiescent-interval slice-selective, Angiography

Introduction

Kidney transplantation is considered to be the treatment of choice for patients with end-stage renal disease. The number of total kidney transplants performed in the USA in 2015 was 18,597,1 with a 1-year survival rate between 80 and 95%.2 Despite continuing improvements, postoperative complications occur in 12–20% of cases3–5 and are due to vascular and nonvascular causes. Vascular causes occur in 3–15% of all transplants.6 Most commonly, these include renal artery stenosis or thrombosis formation in the renal vein, with less common vascular complications that include aneurysm, pseudo-aneurysm, or arteriovenous fistula.7 Renal artery stenosis represents 75% of all vascular complications and can leads to transplant loss.8,9 Early detection and treatment of these complications are crucial to save the transplant and to improve morbidity and mortality.

Invasive digital subtraction angiography (DSA) remains the gold standard for vascular evaluation but is usually reserved for therapy or when the diagnosis cannot be made by noninvasive techniques. Due to its availability and noninvasiveness, color flow Doppler ultrasonography (CFDU) is the most widely performed examination for the initial evaluation of patients suspected of having post-transplant vascular complications.5 However, this technique only estimates the degree of renal artery stenosis indirectly based on arterial flow velocity measurements, and therefore, is dependent on the visualization of the entire donor renal artery and the surgical anastomosis. Furthermore, ultrasound evaluation has often been noted to be limited by operator dependency, it is more difficult to perform and interpret in obese patients, and there is an ongoing debate on the appropriate thresholds for Doppler parameters.10–13 Other noninvasive imaging techniques include contrast-enhanced computed tomography angiography (CTA) and contrast-enhanced magnetic resonance angiography (CE-MRA). These noninvasive angiography techniques directly measure the size of the stenotic artery or identify venous occlusions, the use of contrast agents in post-renal transplant patients has raised concerns because of studies that demonstrated the risk of nephrotoxicity due to iodine-based contrast agents in CTA and nephrogenic systemic fibrosis (NSF) due to gadolinium-based contrast agents in CE-MRA, especially if the glomerular filtration rate (GFR) is <30 mL/m/1.73 m2.14 Recently, several non-contrast enhanced MRA (NCE-MRA) techniques have emerged, which provide a potential alternative. At Northwestern University, NCE-MRA is often used in post-renal transplant patients for a more detailed evaluation of vascular lesions found on CFDU or for definitive diagnosis when the CFDU is inconclusive in patients that have a strong clinical suspicion of vascular lesions. The NCE-MRA protocol (Table 1) includes two techniques that have been optimized for transplant vascular evaluation: quiescent-interval slice-selective (QISS) MRA15 and inversion-prepared inflow-dependent three-dimensional (3D) balanced steady-state free precession (3D bSSFP) MRA.16

Table 1

Typical imaging parameters for SSFP MRA and QISS MRAs

NC-MRA sequenceSSFP MRAQISS
Matrix256/0/0/192400/0/0/195
FOV (mm)320400
TE (ms)1.91.60
TR (ms)3.8640.04
Section thickness (mm)1.03.0
Flip angle9090
Bandwidth (Hz)781658
ExcitationSlab-selectiveSlice-Selective
Mean acquisition time3 m5 m (approximately 0.3 s/slice)
Voxel size (mm)1.3 × 1.3 × 2.00.5 × 0.5 × 3.0

This study aimed to evaluate the image quality and diagnostic performance of NCE-MRA for the evaluation of renal transplant vasculature. The results were compared with the DSA as a reference standard when available.

Material and methods

Data collection

This HIPAA compliant retrospective study was approved by the institutional review board at Northwestern University and the requirement for written informed consent was waived. At a single academic medical center, a PACS search was performed to identify all patients that underwent NCE-MRA (QISS and 3D bSSFP) during the last 4 years for the evaluation of clinical suspicion of transplant renal artery stenosis. Clinical information that included the transplant date, GFR at the time of the exam, CFDU reports and the long-term GFR were recorded. For patients that subsequently underwent DSA for further evaluation of the allograft vasculature, angiography images were reviewed, and the exam results were collected.

Study cohort

A total of 43 patients (mean age: 56 years (range 33–76; 30 males)) were included in this study. The time from the renal transplant to the NCE-MRA date was from 10 days to 15 years (average 2.3 years). All patients were referred for clinical, or ultrasound, or both on suspicion of renal artery stenosis at the anastomosis site. In total, 41 patients (95%) underwent CFDU of the allograft vasculature at our institution before referral for MR imaging. CFDU, which was available for 41 patients, demonstrated elevated arterial velocity at the anastomosis in 34 patients (83%), a parvus tardus waveform in the arcuate arteries that was suspicious for stenosis in two patients (5%), decreased velocity and resistivity indices at the transplant artery suggested iliac stenosis in one patient (2%), and a normal CFDU but delayed graft function in four patients (10%). Two patients did not undergo a CFDU at our institution but were referred for MRA due to strong clinical suspicion (one patient) or abnormal renal scintigraphy (one patient). After the NCE-MRA examination, 12 patients (28%) underwent subsequent renal transplant angiography for further evaluation and possible treatment.

Magnetic resonance imaging

All NCE-MRA exams were performed on 1.5T scanners (MAGNETOM Avanto; Siemens Medical Solutions, Erlangen, Germany) equipped with a 12-element body phased-array coil. The imaging protocol included localizer images of the pelvis that used standard two-dimensional (2D) bSSFP sequences in axial and coronal reformats to identify the transplanted and ipsilateral iliac vessels followed by NCE-MRA sequences (Table 1).

For 3D bSSFP MRA, the area of interest was localized by using scout images. A 150 mm-thick inversion RF pulse was applied from the top of the axial 3D slab inferiorly to suppress the signal from veins and static tissue, with an inversion time of 800–1,200 ms. The images were acquired during free-breathing without respiratory compensation and with electrocardiogram (ECG) gating. The typical scan parameters for the 3D bSSFP MRA sequence were as follows: repetition time 3.8 ms, echo time 1.9 ms; flip angle, 90°; voxel size, 1.3 × 1.3 × 2.0 mm interpolated to 0.65 × 0.65 × 1.0 mm; and field of view, 280 × 320 mm. Total scan time was approximately 3 m (1.5–3.5 m) depending on the number of acquired imaging sections and the inversion time.

The QISS MRA was acquired in an axial plane from the level of the abdominal aorta to the level of the common femoral arteries bifurcation. The QISS sequence is a 2D ECG-gated technique that relies on the application of a slice-selective saturation radiofrequency (RF) pulse to suppress the static tissue, a tracking venous saturation RF pulse to suppress venous signal, and after a quiescent-interval of approximately 230 ms, a chemical shift-selective fat saturation RF pulse and readout are applied to image the arterial inflow. Typical scan parameters for the QISS MRA sequence included: 70° flip angle, TR = 4 ms, TE = 2 ms, TI = 1,200 ms, FOV = 36 × 36 × 24 cm, and BW = ± 75 kHz with a 256 × 256 matrix for a true spatial resolution of 1.4 × 1.4 × 2 mm3 resolution interpolated through zero-filling to 0.7 × 0.7 × 1.0 mm3 (total scan time was approximately 5 m depending on the area covered and the heart rate).

Image analysis

The NCE-MRA images were evaluated separately by two radiologists with 5 years of experience in cardiovascular imaging. Thin multiplanar reformats and maximum intensity projection images that were derived from the 3D bSSFP and QISS MRA were used to assess image quality and stenosis grade (Fig. 1). Since the patients were referred for suspected transplant renal artery stenosis, the analysis focused on the region of anastomosis of the transplanted renal artery to the external iliac artery. For patients with an accessory transplant artery, this vessel was noted but the larger transplant artery was used for qualitative scoring. The presence of any other concomitant pathology was noted, especially arterial or venous thrombosis.

Normal QISS MRA and reformat in a 48-year-old male with a 1-year history of right lower quadrant kidney transplant that presented with mild renal dysfunction (GFR = 48) and abnormal Doppler ultrasound (a and b); Normal appearance of a 3D SSFP MRA reformat in a 60-year-old female with a 3-year history of right lower quadrant kidney transplant presented with renal dysfunction (GFR = 32) and abnormal Doppler ultrasound (c and d).
Fig. 1  Normal QISS MRA and reformat in a 48-year-old male with a 1-year history of right lower quadrant kidney transplant that presented with mild renal dysfunction (GFR = 48) and abnormal Doppler ultrasound (a and b); Normal appearance of a 3D SSFP MRA reformat in a 60-year-old female with a 3-year history of right lower quadrant kidney transplant presented with renal dysfunction (GFR = 32) and abnormal Doppler ultrasound (c and d).

No further vascular evaluation was obtained in these patients. Flow contrast improved, and motion sensitivity reduced with QISS; however, the quality of multiplanar reformats was better with 3D SSFP due to the reduced slice thickness. SSFP, steady-state free precession; QISS, quiescent-interval slice-selective; MRI, magnetic resonance imaging; SSFP, Steady-state free precession; QISS, quiescent-interval slice-selective; GFR, glomerular filtration rate; DSA, digital subtraction angiography.

The images were rated for the image quality and the confidence in transplant arterial anastomosis interpretation, and for the degree of stenosis using a four-point Likert scale as follows: (1) overall image quality and the confidence in transplant arterial anastomosis interpretation, 0 = severe artifact or not interpretable (any type of artifact that degraded the images with no visualization of the anastomosis region or no signal in the region), 1 = moderate artifact but interpretable (presence of artifact but the anastomosis can be evaluated), 2 = good image quality, mild artifact (mild artifact in the image but the anastomosis region was only mildly affected), and 3 = excellent image quality, no artifact (perfect, homogeneous image without any type of artifact); (2) the degree of stenosis at the anastomosis region: 0 = no stenosis, 1 = mild stenosis estimated 0–50%, 2 = moderate stenosis estimated 51–75%, and 3 = severe stenosis estimated at >75%.

Statistical analysis

Interobserver agreement for image quality and qualitative grading of stenosis between both readers and agreement in grading image quality and qualitative stenosis grading scores between the magnetic resonance imaging (MRI) sequences for each reader were analyzed using weighted kappa test (κ). Kappa interpretation was based on Altman guidelines as follows: κ<0.2, poor agreement; κ = 0.21–0.40, fair agreement; κ = 0.41–0.60, moderate agreement; κ = 0.61–0.80, good agreement; κ = 0.81–1, very good agreement. Cohen’s kappa (κ) was computed using SPSS (IBM Corp. Released 2013. IBM SPSS Statistics for Windows, Version 22.0. Armonk, NY: IBM Corp).

Results

The clinical results are summarized in Table 2 and scoring for both readers are summarized in Table 3.

Table 2

Clinical data summary of the cohort including demographics, renal function (GFR) at the time of NC-MRA imaging, peak Doppler velocity of the transplant renal artery at the anastomosis, and resistive indices of the arcuate arteries

PatientGenderAge (years)Time from transplant (day)GFR at time of MRIDoppler velocity (cm/s)RI at arcuate arteries
1M615,01837174elevated
2F4023927313normal
3F761053216elevated
4M633629308elevated
5M199853440normal
6M604336200elevated
7F624,47337126parvus tardus
8F555,2812668low RI
9F652527156elevated
10M6014232203elevated
11M632,33438294normal
12F652,39017221elevated
13M4725025321normal
14F7611359487normal
15M642,26236334elevated
16M5040940328normal
17F6619737436normal
18M717243458normal
19F197045262normal
20M59975142low RI
21M6343330157elevated
22M5412731128parvus tardus
23M514728236normal
24M38212>60347normal
25F3319122404low RI
26M585542309elevated
27M507432322normal
28M493939290normal
29M593,66236218normal
30M554,44937280elevated
31M427736448normal
32M461135NANA
33M7110532358normal
34M4463835NANA
35M698142237normal
36M6145432203elevated
37F58510372elevated
38M4948033412normal
39M6615836412elevated
40M576816425normal
41F686927393normal
42F588733287normal
43M5530940318elevated
Table 3

There was good inter-rater agreement on qualitative scores for both reviewers

Overall image quality and confidence for anastomosis evaluationReader 1 QISSReader 2 QISSReader 1 bSSFPReader 2 bSSFP
00 (0%)0 (0%)5 (12%)5 (12%)
10 (0%)1 (2%)2 (4%)4 (9%)
218 (42%)23 (54%)18 (42%)16 (37%)
325 (58%)19 (44%)18 (42%)18 (42%)
Stenosis evaluationReader 1 QISS (N = 43)Reader 2 QISS (N = 43)Reader 1 bSSFP (N = 38)Reader 2 bSSFP (N = 38)
017 (40%)18 (42%)17 (45%)16 (42%)
110 (23%)7 (16%)8 (21%)9 (24%)
24 (9%)9 (21%)4 (10%)5 (13%)
312 (28%)9 (21%)9 (24%)8 (21%)

For overall image quality and confidence in anastomosis interpretation, all QISS images were scored as good or excellent (mild or no artifact) by reader 1 (scores 2 and 3). Reader 2 scored all images as good or excellent (scores 2 and 3) except for one study that was scored as 1 (moderate artifact). There was fair agreement (κ = 0.507, p<0.001). For 3D bSSFP MRA, reader 1 scored five studies (12%) as a severe artifact or not interpretable (score 0), two studies (4%) as a moderate artifact (score 1), and 36 studies (84%) as good or excellent (scores 2 and 3). Reader 2 scored five studies (12%) as a severe artifact or not interpretable (score 0), four studies (9%) as a moderate artifact (score 1), and 34 studies (79%) as good or excellent (scores 2 and 3). There was moderate inter-rater agreement (κ = 0.755, p<0.001).

For the degree of stenosis evaluation, for the QISS images, reader 1 rated all 43 studies for stenosis as follows: 17 studies (40%) demonstrated no evidence of stenosis, 10 studies (23%) demonstrated mild stenosis, four studies (9%) demonstrated moderate stenosis, and 12 studies (28%) demonstrated severe stenosis. Reader 2 rated all 43 studies for stenosis as follows: 18 studies (42%) demonstrated no evidence of stenosis, seven studies (16%) demonstrated mild stenosis, nine studies (21%) demonstrated moderate stenosis, and nine studies (21%) demonstrated severe stenosis. There was moderate inter-rater agreement (κ = 0.741, p<0.001).

For 3D bSSFP MRA, both readers rated 38 studies (not interpretable anastomosis in five studies for both readers) as follows: Reader 1: 17 studies (45%) demonstrated no evidence of stenosis, eight studies (21%) demonstrated mild stenosis, four studies (10%) demonstrated moderate stenosis, and nine studies (24%) demonstrated severe stenosis. Reader 2 rated 38 out of the 43 studies for stenosis as follows: 16 studies (42%) demonstrated no evidence of stenosis, nine studies (24%) demonstrated mild stenosis, five studies (13%) demonstrated moderate stenosis, and eight studies (21%) demonstrated severe stenosis. There was moderate inter-rater agreement (κ = 0.766, p<0.001).

Both techniques were compared in the evaluation of the anastomosis stenosis for both readers and there was a moderate agreement for reader 1 and a fair agreement for reader 2 (κ = 0.653 and 0.432, respectively, p<0.001).

Correlation with DSA: DSA was performed in 12 patients for further evaluation of the renal transplant vasculature. The results are summarized in Table 4. Out of 12 patients, 10 patients had strong imaging evidence on NCE-MRA of either post-transplant renal artery anastomosis, external iliac artery at the anastomosis, or the common iliac artery at the anastomosis site (score 2 and 3 for stenosis). One patient was referred for evaluation of a renal intra-parenchymal arteriovenous malformation that was observed on CFDU and in one patient the NCE-MRA demonstrated mild anastomotic stenosis (score 1) for both readers and underwent DSA for clinical suspicion of anastomosis significant stenosis, DSA demonstrated mild stenosis without increased pressure gradient along the stenosis and was not treated.

Table 4

Comparison between DSA and NC-MRA sequences

PatientDSA resultQISS score reader 1QISS score reader 2bSSFP score reader 1bSSFP score reader 2Pretreatment GFRPost-treatment GFR
2333332738
42a33Not interpretableNot interpretable2936
83a33Not interpretableNot interpretable26>60
13b00000No treatment of anastomosis
173333237>60
182333343>60
2033333545
283333339>60
3511111No treatment of anastomosis
39222223656
40222221623
41333332737

DSA demonstrated the presence of renal transplant artery stenosis at anastomosis in eight patients (Fig. 2), external iliac artery stenosis at the level of the renal transplant anastomosis in one patient, common iliac artery stenosis in one patient and intra-parenchymal arteriovenous fistula in one patient, not seen on NCE-MRA but seen on CFDU, this patient had normal anastomosis on both imaging modalities. One patient underwent DSA for clinical and CFDU suspicion of stenosis at the anastomosis despite only mild stenosis on QISS MRA (score 1 for both readers), subsequently, the DSA demonstrated only mild stenosis. In a patient with severe stenosis of the common iliac artery that resulted in the loss of distal intravascular signal in the transplant renal artery with 3D bSSFP, whereas the lesion was successfully visualized on QISS MRA (Fig. 3) and confirmed on DSA. Another case of severe stenosis of the external iliac artery resulted in a nondiagnostic 3D bSSFP sequence but was successfully depicted on QISS MRA and confirmed on DSA.

59-year-old male with right lower quadrant renal transplant presented 2 months after transplantation for evaluation high velocity at the artery anastomosis and renal dysfunction (GFR = 5).
Fig. 2  59-year-old male with right lower quadrant renal transplant presented 2 months after transplantation for evaluation high velocity at the artery anastomosis and renal dysfunction (GFR = 5).

Axial and reformats QISS MRA (a, b, and c) demonstrate obliteration of the transplant renal artery in its post anastomotic segment (arrows). The anastomosis itself is patent (arrow). 3D SSFP MRA demonstrates similar findings (arrow in (d)). The patient underwent DSA exam (e and f) that demonstrate irregularity and high-grade post anastomotic stenosis of the renal artery (arrows) treated with balloon angioplasty and stenting with improvement of the renal function post-treatment. SSFP, Steady-state free precession; QISS, quiescent-interval slice-selective; GFR, glomerular filtration rate; DSA, digital subtraction angiography.

55-year-old female with left lower quadrant renal transplant in 2003 presented with transplant dysfunction (GFR = 26).
Fig. 3  55-year-old female with left lower quadrant renal transplant in 2003 presented with transplant dysfunction (GFR = 26).

Doppler ultrasound demonstrated low velocity and resistivity index at the renal artery anastomosis concerning the proximal stenosis. Coronal reformat of QISS MRA (a) demonstrates severe stenosis of the left common iliac artery proximal to the renal transplant anastomosis (arrow). The renal anastomosis was normal on sagittal reformat of QISS MRA (b). 3D SSFP MRA was not diagnostic due to lack of signal. Patient underwent subsequent DSA (c) that showed the left common iliac stenosis (arrow) and post-stenting appearance. Of note, there was excellent correlation with the QISS MRA. The kidney function recovered post-treatment. SSFP, Steady-state free precession; QISS, quiescent-interval slice-selective; GFR, glomerular filtration rate; DSA, digital subtraction angiography.

Discussion

NC-MRA techniques are widely used due to being risk-free, which is related to the use of gadolinium or iodine-based agents. Especially in patients with renal transplant and low GFR, clinicians tend to avoid the use of iodine contrast due to nephrotoxic effects, especially in patients with a GFR of <30. The use of a gadolinium-based agent is limited by the risk of NSF; however, this risk is low with the newer agents, but there is a newly emerging risk of the deposition of contrast in the brain that has unclear long-term significance. The QISS MRA technique is predominantly used to evaluate suspected lower extremity peripheral arterial disease.15,17,18 To the best of our knowledge, this is the first study to investigate the use of QISS MRA in transplant renal artery assessment. QISS MRA reliably depicted post-transplant renal arterial anatomy. In addition, this study demonstrated that the combination of both NCE-MRA techniques (QISS and 3D bSSFP) could evaluate the transplant artery anastomosis for stenosis in all patients that were suspected of transplant artery stenosis with a good-to-excellent inter-rater agreement. In a small cohort of patients where DSA correlation was available, NCE-MRA accurately graded transplant anastomotic stenosis comparison with the reference standard. Both NCE techniques proved complementary: however, QISS provided diagnostic image quality in a larger proportion of the subjects, the 3D bSSFP acquisition afforded thinner slices for improved quality of multiplanar reconstructions.

Several previous studies demonstrated that 3D bSSFP-based NCE-MRA sequences could evaluate renal transplant arteries.8,16,19–21 In this study, adequate evaluation of transplant artery stenosis using 3D bSSFP MRA was feasible in 38 out of 43 cases (88%). Previous studies have reported slightly higher numbers, which included 87% in a study of 15 patients,16 and 95.4% of cases in a larger study of 369 examinations.19 The differences in the reported success rate could be related to differences in the study design, technique implementation, or patient populations.

The 3D bSSFP sequence has a scan time of few minutes, compared with approximately ⅓ s for each 2D QISS slice. Consequently, the 3D bSSFP sequence had an increased risk of artifacts from respiration and bowel motion. In addition, suboptimal positioning of the in-plane inversion slab could result in saturation of the vessels of interest. Technical failure of the 3D bSSFP MRA sequence occurred in five cases in this study, further evaluation of these cases demonstrated that in two cases, the stenosis was at the common iliac and external iliac arteries proximal to the anastomosis that led to the loss of distal flow secondary to severe stenosis that was successfully depicted on the QISS images. In another three cases, there was a technical error in positioning the inversion slab that resulted in inflowing arterial blood signal suppression as the result of this error. The combination of both sequences could potentially provide more confidence in the evaluation the arterial anastomosis. In addition, one sequence could serve as a backup when the other was limited due to artifacts or technical limitations.

This study has limitations due to the small number of subjects, the retrospective nature of data acquisition, and the presence of the gold standard reference (DSA) in a small portion of subjects.

Future directions

Non-contrast MRI plays an important role in daily clinical practice. Due to the continuous development of robust non-contrast sequences, this allows for accurate imaging diagnosis, which spares the patient from the injection of contrast material. This study demonstrated the possibility of accurate arterial renal transplant evaluation using a NC-MRA technique, QISS, and confirmed the usefulness of 3D IR bSSFP MRA for this type of indication. This study is the first to report the use of the QISS sequence for this indication, to the best of the authors’ knowledge, and could provide a reference for future research into the use of QISS in the evaluation of the vascular system in renal transplants for a larger cohort and the evaluation of other visceral arteries.

Conclusions

QISS and 3D SSFP showed good inter-rater agreement for image quality and stenosis grade, with more cases being of adequate image quality using QISS. In addition, there was an excellent correlation with the angiography results when DSA was performed on a subset of patients. Further study is required; however, MRA that used a combination of these NCE techniques has potential as a risk-free alternative to CTA and CE-MRA for the evaluation of renal transplant arterial anatomy.

Abbreviations

bSSFP: 

balanced steady-state free precession

CTA: 

computed tomography angiography

DSA: 

digital subtraction angiography

GFR: 

glomerular filtration rate

QISS: 

quiescent-interval slice-selective

NCE-MRA: 

non-contrast enhanced magnetic resonance angiography

NSF: 

nephrogenic systemic fibrosis

Declarations

Acknowledgement

None.

Data sharing statement

Technical appendix, statistical codes, and dataset are available from the corresponding author at [email protected]

Funding

This study was supported by the National Institutes of Health (R01 HL130093 and R21 HL126015).

Conflict of interest

Robert R. Edelman received research support and royalties from Siemens Healthcare. The other authors report no conflicts of interest related to this publication.

Authors’ contributions

Study design (AS, PA, RRE, RA), analysis and interpretation of data (AS, PA), statistical analysis (AS, PA, RRE, RA), manuscript writing, critical revision (AS, PA, MS, AP, PL, JC, RRE, RA). All authors have made a significant contribution to this study and have approved the final manuscript.

References

  1. Hart A, Smith JM, Skeans MA, Gustafson SK, Stewart DE, Cherikh WS, et al. OPTN/SRTR 2015 Annual Data Report: Kidney. Am J Transplant 2017;17(Suppl 1):21-116 View Article PubMed/NCBI
  2. Cecka JM, Terasaki PI. The UNOS Scientific Renal Transplant Registry. Clin Transpl 1992:1-16 View Article PubMed/NCBI
  3. Orons PD, Zajko AB. Angiography and interventional aspects of renal transplantation. Radiol Clin North Am 1995;33(3):461-471 View Article PubMed/NCBI
  4. Shoskes DA, Hanbury D, Cranston D, Morris PJ. Urological complications in 1000 consecutive renal transplant recipients. J Urol 1995;153(1):18-21 View Article PubMed/NCBI
  5. Brown ED, Chen MY, Wolfman NT, Ott DJ, Watson NE. Complications of renal transplantation: evaluation with US and radionuclide imaging. Radiographics 2000;20(3):607-622 View Article PubMed/NCBI
  6. Kobayashi K, Censullo ML, Rossman LL, Kyriakides PN, Kahan BD, Cohen AM. Interventional radiologic management of renal transplant dysfunction: indications, limitations, and technical considerations. Radiographics 2007;27(4):1109-1130 View Article PubMed/NCBI
  7. Reyna-Sepúlveda F, Ponce-Escobedo A, Guevara-Charles A, Escobedo-Villarreal M, Pérez-Rodríguez E, Muñoz-Maldonado G, et al. Outcomes and Surgical Complications in Kidney Transplantation. Int J Organ Transplant Med 2017;8(2):78-84 View Article PubMed/NCBI
  8. Tang H, Wang Z, Wang L, Hu X, Wang Q, Li Z, et al. Depiction of transplant renal vascular anatomy and complications: unenhanced MR angiography by using spatial labeling with multiple inversion pulses. Radiology 2014;271(3):879-887 View Article PubMed/NCBI
  9. Bruno S, Remuzzi G, Ruggenenti P. Transplant renal artery stenosis. J Am Soc Nephrol 2004;15(1):134-141 View Article PubMed/NCBI
  10. Pan FS, Liu M, Luo J, Tian WS, Liang JY, Xu M, et al. Transplant renal artery stenosis: Evaluation with contrast-enhanced ultrasound. Eur J Radiol 2017;90:42-49 View Article PubMed/NCBI
  11. Granata A, Clementi S, Londrino F, Romano G, Veroux M, Fiorini F, et al. Renal transplant vascular complications: the role of Doppler ultrasound. J Ultrasound 2014;18(2):101-107 View Article PubMed/NCBI
  12. Ngo AT, Markar SR, De Lijster MS, Duncan N, Taube D, Hamady MS. A systematic review of outcomes following percutaneous transluminal angioplasty and stenting in the treatment of transplant renal artery stenosis. Cardiovasc Intervent Radiol 2015;38(6):1573-1588 View Article PubMed/NCBI
  13. Fananapazir G, McGahan JP, Corwin MT, Stewart SL, Vu CT, Wright L, et al. Screening for Transplant Renal Artery Stenosis: Ultrasound-Based Stenosis Probability Stratification. AJR Am J Roentgenol 2017;209(5):1064-1073 View Article PubMed/NCBI
  14. Dekkers IA, Roos R, van der Molen AJ. Gadolinium retention after administration of contrast agents based on linear chelators and the recommendations of the European Medicines Agency. Eur Radiol 2018;28(4):1579-1584 View Article PubMed/NCBI
  15. Edelman RR, Sheehan JJ, Dunkle E, Schindler N, Carr J, Koktzoglou I. Quiescent-interval single-shot unenhanced magnetic resonance angiography of peripheral vascular disease: technical considerations and clinical feasibility. Magn Reson Med 2010;63(4):951-958 View Article PubMed/NCBI
  16. Liu X, Berg N, Sheehan J, Bi X, Weale P, Jerecic R, et al. Renal transplant: nonenhanced renal MR angiography with magnetization-prepared steady-state free precession. Radiology 2009;251(2):535-542 View Article PubMed/NCBI
  17. Edelman RR, Giri S, Dunkle E, Galizia M, Amin P, Koktzoglou I. Quiescent-inflow single-shot magnetic resonance angiography using a highly undersampled radial k-space trajectory. Magn Reson Med 2013;70(6):1662-1668 View Article PubMed/NCBI
  18. Serhal A, Koktzoglou I, Edelman RR. Feasibility of Image Fusion for Concurrent MRI Evaluation of Vessel Lumen and Vascular Calcifications in Peripheral Arterial Disease. AJR Am J Roentgenol 2019;212(4):914-918 View Article PubMed/NCBI
  19. Zhang LJ, Peng J, Wen J, Schoepf UJ, Varga-Szemes A, Griffith LP, et al. Non-contrast-enhanced magnetic resonance angiography: a reliable clinical tool for evaluating transplant renal artery stenosis. Eur Radiol 2018;28(10):4195-4204 View Article PubMed/NCBI
  20. Bultman EM, Klaers J, Johnson KM, François CJ, Schiebler ML, Reeder SB, et al. Non-contrastenhanced 3D SSFP MRA of the renal allograft vasculature: a comparison between radial linear combination and Cartesian inflowweighted acquisitions. Magn Reson Imaging 2014;32(2):190-195 View Article PubMed/NCBI
  21. Lanzman RS, Voiculescu A, Walther C, Ringelstein A, Bi X, Schmitt P, et al. ECG-gated nonenhanced 3D steady-state free precession MR angiography in assessment of transplant renal arteries: comparison with DSA. Radiology 2009;252(3):914-921 View Article PubMed/NCBI