v
Search
Advanced Search

Publications > Journals > Exploratory Research and Hypothesis in Medicine > Article Full Text

  • OPEN ACCESS

Reducing False Alarms in Intensive Care Units: A Scoping Review

  • Jiayan Huo1,
  • Shufen Wung2,3,
  • Janet Roveda1,3,4 and
  • Ao Li3,4,* 
 Author information  Cite
Exploratory Research and Hypothesis in Medicine   2023;8(1):57-64

doi: 10.14218/ERHM.2022.00026

Abstract

Background and objectives

Excessive false alarms in intensive care units (ICU) cause noise disturbance to patients and develop alarm fatigue among nurses, leading to safety concerns and decreased patient care quality. Evidence-based false alarm reduction strategies are urgently needed in the day-to-day clinical practice. This review aims to synthesize two main human-technology approaches to reduce false alarms generated by the physiologic monitor: customization of alarm settings by nurses and alarm algorithms.

Methods

A broad search was performed using four electronic databases, PubMed, Scopus, EMBASE, and Cumulative Index to Nursing and Allied Health Literature. This review included twenty-eight full-text journal articles focused on both customizations of alarm settings and alarm algorithm improvement for false alarm reduction in the ICU.

Results

Clinical customizations of alarm settings on bedside physiological monitors helps to reduce excessive false alarms. Colleagues also developed alarm algorithms to reduce false alarms in the ICU and achieved excellent performance.

Conclusions

This review suggests collaboration between nurses and engineers to optimize personalized machine learning algorithms has the great potential for false alarm reduction in the ICU.

Keywords

Intensive care units, Clinical alarm, Artificial intelligence, Machine learning, Deep learning

Introduction

Excessive false clinical alarms compromise patient care and safety in intensive care units (ICU). Critical care nurses identified alarms from cardiopulmonary physiologic monitors are one of the most helpful in the ICU, and yet the monitors also contribute to the highest numbers of false alarms,1,2 leading to the sensory overload at work. The bedside physiologic monitors have embedded alarm systems to alert clinicians to life-threatening and imminent changes in a patient’s condition or device malfunction.3 However, it is well known that up to 94% of alarms were false positives or clinically irrelevant.4–7 These unnecessary false or irrelevant alarms have reached a noise peak of over 80 dB creating a noisy and annoying environment for both nurses and patients.8,9 Noise pollution from false alarms is viewed to be the most stressful noise in the ICU.10 It hinders patient recovery and quality of care. Moreover, constant demand and mistrust of the alarm system reduce the alertness of the clinicians leading to alarm fatigue. Alarm fatigue has resulted in desensitization to alarms which compromise patient safety,11 including deaths, permanent loss of function, and unexpected additional care or extended stay. A high number of false alarms remained an unresolved issue in clinical practice.12

There are three categories of false alarms – clinical false alarms, technical false alarms, and false alarms through interventions.2 Clinical false alarms refer to the situation in which the physiological signal exceeds a preset threshold but is not clinically relevant. A default alarm setting in physiologic monitors may not appropriately reflect the individual patient’s baseline or change in condition. For instance, a slow heart rate below 60 triggers an alarm which may be normal physiologically for some individuals or associated with an effective pharmacologic response.13 These high/low heart rate limits should be individually programmed as the same thresholds may not be appropriate for patients in cardiac ICU and those in medical ICU. Thus, customization of alarm settings based on individualized patient status can reduce the number of clinical false alarms and make the alarm more relevant to patient care.14 Graham and Cevach reported that providing nurses training about monitoring systems and regular assessment and individualization of alarm parameters can reduce false critical physiologic monitor alarms.15 Ruppel et al. also pointed out that customizing alarm settings enhances alarms’ clinical relevance; however, alarm customization is a complex process, and little is known about the challenges nurses face when customizing alarms.14 In a mixed-methods study, Ruppel et al. reported variation in nurses’ customization practices and confidence, unit-specific differences in alarm customization, and nurses were frustrated when they could not figure out how to customize specific arrhythmia alarms.14

Technical false alarms refer to the situation corresponding to a variable unrelated to surpassing the preset thresholds. Examples of these technical false alarms include a false asystole crisis alarm due to a low QRS amplitude and motion artifacts. The artifacts had an unfavorable effect on the physiologic waveforms that reduced the signal-to-noise ratio, leading to technical false alarms. Cho et al. found that 45.1% of the patient monitor alarms are false because of technical issues.16 Takla et al. reviewed extensive sources of common artifacts that affect physiologic monitor data,17 including movement artifacts (e.g., patient movement and surgical preparation) or sensor artifacts (e.g., circuit malfunction and losing contact). The detection and filtering of these artifacts can minimize technical false alarms.17–20

Interventions to manage alarms can be a technological implementation within the monitoring system.21 Intelligent alarm algorithms, including machine learning, have attracted full attention for false alarm suppression in the ICU owing to their ability to discover unknown but ponderable information from a massive amount of medical data.22–26 Promising machine learning algorithms can overcome the challenges caused by irregular and high-dimensional ICU data.27 These algorithms allow the integration of data from different sources and the incorporation of background knowledge in the analysis.28 The renowned critical care databases, such as MIMIC-II29 and ANZICS APD,30 offer a wealth of data for machine learning models to learn features and mine the hidden information in the data. Progress has been made in algorithm development and improvement to identify false alarms generated by physiologic monitors and to discriminate feature patterns between alarms due to true patient instability or artifacts.19

The purpose of this review is to synthesize two main human-technology approaches to reduce false alarms generated by the physiologic monitor: customization of alarm settings by nurses and alarm algorithms.

Materials and methods

Data sources and search strategy

This review was performed according to the Guidance for conducting systematic scoping reviews.31 A broad search was performed for reports on the reduction of false alarms from physiologic monitors in ICU using two electronic databases, PubMed and Scopus. The following keywords were used alone or in combination to conduct this search: “ICU”, “false alarm”, “reduction”, “critical care”, “nurses”, “nursing”, “customization”, “alarm fatigue”, “machine learning”, “deep learning” and the search was limited to articles published in English. All articles with publication dates ranging from database establishment to December 2020 were included. Another search was carried out to include non-duplicated articles from the EMBASE and the Cumulative Index to Nursing and Allied Health Literature (CINAHL) databases. To identify further eligible studies, reference lists of the retrieved articles were also examined.

Inclusion and exclusion criteria

All studies that reduced the false alarms from physiologic monitors in ICU and employed participants or biomedical databases were considered eligible for our study. Specifically, eligible studies that met the following inclusion criteria were included: (1) clinical setting - ICU; (2) outcomes - reduction, causes, or consequences of the false alarms in the ICU, and (3) original research articles – reports of studies written by the researchers performed the studies. Studies were excluded according to the following exclusion criteria: (1) studies duplicated earlier publications, (2) studies were not related to false alarms in the ICU setting, (3) outcomes were not related to false alarms causes or reduction, (4) studies contained overlapping methods/algorithm, and (5) meta-analyses or systematic reviews.

Data extraction and appraisal

Two reviewers (JYH and SFW) extracted the following data from eligible studies: author name, publication year, study subjects, sample size, study design, interventions, and outcomes. The reviewers screened articles to ensure the studies met the inclusion criteria. A third reviewer (AL) was consulted in case of disagreement. The studies were appraised from two perspectives in terms of false alarm reduction: customization of alarm settings and alarm algorithm improvement.

Results

Literature selection and characteristics of eligible studies

Figure 1 illustrates a flowchart of literature search and selection. A total of 907 articles were initially identified through the literature search. Of these articles, 419 were from PubMed, 344 from Scopus, 120 from Embase, and 24 from CINAHL. Among these, 176 duplicate articles were removed. After the title and abstract of the remaining articles were screened, 604 additional articles were removed because they were not related to false alarm reduction in the ICU setting. After a full-text review, 99 reports were further excluded. These include reports not involving alarms in the ICU (n = 34), outcomes not related to false alarm causes or reduction (n = 40), articles with overlapping algorithms (n = 22), and reviews (n = 3). Finally, a total of 28 studies were included in this scoping review. The characteristics of the studies included in this review are presented in Tables 1 and 2.3,12,14–16,19,20,22–25,32–48

Flowchart of literature search and study selection for this review.
Fig. 1  Flowchart of literature search and study selection for this review.
Table 1

Customization of alarm settings

StudyParticipantStudy Design/InterventionOutcome
Graham et al.1530 nurses in a 15-bed care unitNurses were trained on the customization of alarm settings for individual patients.Critical monitor alarms were reduced by 43% from baseline.
Ruppel et al.14ECG alarm customization of 298 patients in 3 ICUs and 27 nurses’ customization clinical reasoning was exploredConducted semi-structured interviews to explore the nurses’ customization of clinical reasoning58.7% of patients had alarms customized and factors that affect customizations were revealed.
Wung et al.3ICU nursesData analysis through a structured interviewIdentified alarms from physiological monitors that contribute most to sensory overload.
Cho et al.16ICU nurses and devices from 48 critically ill patient casesData analysis through observation of alarms in 48 hoursNurses personally set an alarm range that reflected the patients’ condition in no more than 9 cases (18.8%).
Ruppel et al.3327 ICU nursesThematic analysis of semi-structured interviewsEstablished a conceptual model to reveal the features affecting the customization of alarm.
Wung et al.3516 ICU nursesAnalysis of decision-making process via semi-structure interviewsIdentified the factors that nurses considered when deciding the response to an alarm in ICU
Honan et al.36406 nurses’ comments on perceptions of clinical alarmsComments analysis using the Krippendorff methodBedside nurses should set limits appropriate to the individual patient. Improved technology that recognizes trends in values to decrease nuisance alarms.
Fidler et al.3777-bed from 5 ICUsQuantitative analysis of heart rate alarmsHeart rate parameter adjustment did not lead to a reduction in alarms (p = 0.57).
Lewis et al.3836-bed ICU/SDUCEASE BundleThe number of auditory monitor alarms decreased by 30.45% without adverse patient events.
Bi et al.3293 ICU clinical nursesThe experimental group (n = 47) received alarm management trainingThe number of total alarms and nonactionable alarms from the physiological monitor decreased significantly in the experimental group (p < .001).
Allan et al.3418-bed cardiac and vascular surgery ICUTested a bundle set of published practices, including education to staff regarding alarm customization, to reduce monitor alarmA bundle set of alarm reduction strategies decreased the average alarms per monitored bed by 61%.
Table 2

Algorithms for false alarm reduction

StudyDataAlgorithmMethodOutcome
Mäkivirta et al.39Patients’ data from a multi-parameter monitorStatisticalMedian filterThe true alarms rate increased from 12% to 49% without missing correct alarms.
Zong et al.40PhysioNet’s MIMICStatisticalArterial blood pressure waveform quality combined with PPG waveformDeveloped an algorithm that can suppress 100% of false ECG arrhythmia alarms based on the dataset.
Li et al.41MIMIC IIStatisticalSignal quality indices fusion and Kalman filterAn accurate heart rate estimate was achieved in the presence of persistent noise and artifacts.
Zong et al.42MIMICStatisticalSignal quality assessment and fuzzy logic98.2% (159/162) of false arterial blood pressure alarms were rejected while accepting 99.8% (441/442) of true alarms.
Fernandes et al.12Monitoring data and vital signs during surgical casesStatisticalAutomatic reasoning systemThe reasoning algorithm filtered and reduced the notifications without compromising patient safety.
Pimentel et al.45PhysioNet Challenge 2014Machine LearningHidden Markov model and signal quality indexThe overall score of the algorithm for the third phase of Physionet Challenge 2014 reached 83.47%.
Hravnak et al.198-week vital sign data collected in surgical-trauma stepdown-unitMachine LearningLogistic regression for feature selection and multiple machine learning algorithms were employed for classificationThe testing set achieved AUC scores of 0.94 respiratory rate, 0.84 blood pressure, and 0.72 SpO2.
Lameski et al.22MIMIC IIMachine LearningRF, SVM, and ERTThe ERT suppressed over 90% of the TACHY false alarms with a low true alarm suppression rate.
Au-Yeung et al.23PhysioNet Challenge 2015Machine LearningRFAchieved a score of 83.08 in the real-time category on the hidden test set.
Li et al.24MIMIC IIMachine LearningSignal quality index assessment and relevance vector machineFalse alarm suppression results were 86.4% for asystole, 100% for extreme bradycardia, and 27.8% for extreme tachycardia without suppressing true alarm
Eerikainen et al25PhysioNet Challenge 2015Machine LearningSignal pair selection and RFThe algorithm achieved a 93% true positive and 83% true negative rate in classifying cardiac arrhythmia alarms.
Srivastava et al.43PhysioNet Challenge 2015Machine LearningRF and threshold approach assemblingThe overall model performance reached an accuracy of 83.96%.
Krasteva et al.44PhysioNet Challenge 2015, AHA, EDB, SVDB, MIT-BIHMachine LearningSignal quality and Decision treeThe algorithm achieved an overall real-time score of 80%.
Antink et al.46PhysioNet Challenge 2015Machine LearningBinary classification trees, Discriminant analysis classifier, SVMThe algorithm achieved an overall real-time score of 75.55%.
Silva et al.47MIT-BIH, CYBHiDeep LearningCNNThe positive prediction of R-peak was enhanced compared to the Pan-Tompkins algorithm in both databases.
Mousavi et al.48PhysioNet Challenge 2015Deep LearningCNN with a two-step BP trainingThe algorithm achieved a sensitivity of 93.88% and specificity of 92.05% when considering three different signals.
Hooman et al.20PhysioNet Challenge 2015Deep LearningCNN and DFOThe proposed method for training NNs improves the performance of detecting false alarms in ICU compared to backpropagation-trained networks.

Customization of alarm settings

Customization of alarm settings refers to adjusting the alarm settings based on each patient’s condition to minimize the irrelevant alarms without missing essential events. Graham et al.15 conducted a quality improvement project in which the nurses were trained to customize alarm parameters based on each patient’s needs and troubleshoot common monitor problems. Results showed that the number of high priority alarms decreased by 43% (from 16,953 to 9,647) in 18 days, suggesting customizing alarm settings can significantly decrease the number of alarms. In a randomized control study, Bi et al. conducted a 12-week alarm management training that significantly decreased the number of total alarms and nonactionable alarms (p < 0.001).32 However, in a Korean study, Cho and colleagues found that nurses only set personalized alarm ranges that reflected the patient’s conditions in only 18.8% of cases.16 The underlying reasons for many nurses in practice failing to set individualized alarm ranges for patients should be further studied.

It should be noted that the process of alarm customization involves clinical reasoning and decision-making. Some studies have examined which factors play an important role in this process. Ruppel et al.14 found that a nurse’s clinical reasoning was affected by clinical expertise, lack of customization education, and negative experiences. A conceptual model was established to reveal the features that affect the customization of alarm and decision making.33 Allan et al.34 reported a bundled set of alarm reduction strategies, including alarm customization, reduced 61% of average alarms per monitored bed. Wung et al.35 reported that nurses’ responses to the alarms were influenced by the type of alarm, workload status, and training on alarm devices. Multiple interacting factors may influence nurses’ alarm customization practice, however, clinical supports, such as education, structural training, and unit culture and policies, are of great importance to alarm management.

Honan et al. mentioned that clinical alarm management is important but not a panacea.36 Evidence indicates that threshold customization is not enough to result in a significant reduction in alarm rates (p = 0.57).37 Threshold adjustments manually can be either too small or too late to reduce the false alarms. Lewis et al.38 employed a nurse-driven tool, the Communication, Electrodes, Appropriateness, and Setup alarm parameters, and Education (CEASE) bundle, and reduced 30% of the total number of monitoring alarms. This nurse-driven patient-customized monitoring bundle not only decreased the total number of monitoring alarms but also improved nurse perception of alarm fatigue.

Alarm algorithms for false alarm reduction

Artifacts in physiological signals, as a common source of the false alarm, have significantly lowered the patient care quality and safety in the ICU.6 Statistical and machine learning algorithms are common approaches to detect artifacts.2 The algorithms to detect artifacts can be improved from three aspects: signal acquisition, validation of alarms, and alarm generation.1

Several statistical approaches have been employed to improve signal extraction, such as median filters,39 multivariate statistical methods, and signal quality assessment methods.40–42 These statistical manners are widely applied to reduce the artifacts.

The alarm fatigue can also be alleviated by reasoning the alarm generation and validation. Fernandes et al.12 proposed a reasoning algorithm to group a set of alarms. The results showed that up to 99.3% (582/586) of total alarms were reduced without compromising patient safety.

In the past two decades, machine learning has raised the ever-increasing interest in classifying alarm events owing to its compatibility with complex data contexts.18 Machine learning combines statistics and computer science which uses input data to perform a task without being explicitly programmed (i.e., “hardcoded”).49 The parameters/architecture of the model can automatically alter based on observed data. Machine learning can deal with high dimensional data with multiple features as well as complex data formats, such as images, waveforms, etc. There are two types of classification algorithms in machine learning, linear and non-linear. The linear classifiers have the advantage of fast processing speed and the prerequisite is that data can be separated by a hyperplane. The non-linear classifiers, such as support vector machine (SVM) with Gaussian kernel function, random forest, and deep learning, have a higher tolerance to the noise and better performance. In this review, the term “machine learning” refers to traditional machine learning algorithms, including SVM, logistic regression, discriminant analysis classifier, and other tree-based classification algorithms (random forest [RF] and extremely randomized tree [ERT]).

In the PhysioNet/Computing in Cardiology challenge 2105, several novel approaches integrated with machine learning have been proposed for the classification of false and true alarms.50 Eerikainen et al.25 processed three physiological signals to confirm arrhythmias by choosing most matching signals pairs based on F1-score. Then specific arrhythmia features were calculated with customized windows varying from 14–16 seconds, followed by the classification through a random forest model. Srivastava et al.43 ensembled random forests with a threshold approach. Random forest was implemented first for combinations of features extracted from electrocardiogram (ECG), arterial blood pressure, and photoplethysmogram (PPG), followed by a threshold set for parameters based on pulsatile waveforms. Krasteva et al.44 employed a decision tree and introduced ECG quality to increase the signal-to-noise ratio. Then, a short scan interval of 3 to 7.5 seconds made it possible to give immediate feedback on arrhythmia events. Pimentel et al.45 used the hidden Markov model and signal quality index to detect a heartbeat. Signal quality makes a significant difference in the classification of the alarm, even though it is not directly related to the arrhythmia characteristics.23 The algorithm proposed by Antink et al.46 involved multiple types of classifiers. Different classifiers were trained for specific arrhythmia alarms as well as a global classifier for general false alarms detection.

Deep learning, as a subfield of machine learning, has drawn increased attention to detecting or suppressing false alarms in the ICU. Deep learning structures algorithms in multiple layers to create an artificial neural network that mimics human brain structure and can learn from data and make decisions on its own. Deep neural networks can estimate much more complicated decision boundaries due to non-linear combinations of layers, thus, they can benefit from large amounts of data while the performances of traditional machine learning models plateau concerning increasing data size.51

Silva et al.47 proposed a real-time approach based on convolutional neural networks (CNN) to enhance the positive prediction of heartbeat identification compared to the Pan-Tompkins algorithm, even though there was a slight decrease in the sensitivity of true heartbeat identification. Mousavi et al.48 trained the CNN by a two-step backpropagation (BP) algorithm and achieved a sensitivity of 93.88% and a specificity of 92.05% for false alarm classification. Instead of BP, Hooman et al.20 improved the deep neural networks’ performance in the detection of false alarms in the ICU by applying the dispersive files optimization (DFO) algorithm to find an optimal weight for the model.

Discussion

The high rate of false alarms caused by clinical or technical factors significantly burdens the nurse, resulting in alarm fatigue, and compromising patient care quality and safety. This review summarized customization of alarm settings and alarm algorithm improvement as two approaches for false alarm reduction in the ICU.

Default alarm settings are inappropriate to reflect an individual’s condition change and not all signals that exceed the default threshold are clinically relevant. On the other hand, artifacts that have unfavorite impacts on physiologic waveforms or reduce signal-to-noise ratios can cause technical false alarms. Moreover, false monitor alarms can also arise from the interaction of technical and clinical factors. According to these causes of false alarms in ICU, customization of alarm settings and improved alarm algorithms can be great potential solutions to lower false alarms in the ICU.

Customizing alarm settings based on individual conditions can be a promising solution for clinical false alarms. Several studies found adjusting the alarm settings based on each patient’s condition can reduce false alarms.15,16,32 Colleagues also noted the importance of alarm customization involves clinical reasoning and decision making.14,33–35 Previous studies also found that human behaviors play a role in response to alarms.52,53 Experienced nurses can dynamically change their activities according to the received information.53 A nurse-driven patient-customized monitoring bundle can further reduce false alarms and improve nurse perception of alarm fatigue.38 Therefore, Factors that affect nurses’ customization and alarm management suggest that education and good unit culture are essential.

Although customization of alarms can reduce false alarms, it is not adequate to resolve all false alarms in the ICU. Intelligent alarm algorithms can effectively reduce technical false alarms by using multi-dimensional data to train the model. In the past, statistical approaches were widely applied to reduce the artifacts in false alarm reduction.39–42 Colleagues also found that combining information from different sensors can increase the positive predictive value of alarms.54 Recently, machine learning has raised ever-increasing interest in classifying alarm events from multiple physiological signals.25,43–46,50 In addition, in the recent five years, deep learning has drawn increased attention to detecting or suppressing false alarms in the ICU.20,50,51 These studies show that modern machine learning methods with sensor fusion can effectively reduce false alarms.

Limitations

It should be noted that this study has a few limitations. First, this review was restricted to English-language papers published in academic journals, which may result in language and publication bias. Second, most alarm customization studies were conducted in America which may affect the generalizability of findings to other countries. Finally, the majority of the studies utilized AI/statistical approaches to suppress a single type of false alarm. Further investigations applying different technical approaches for multiple types of false alarm reduction are necessary.

Future directions

Human-in-the-loop approaches have great potential to further reduce the false alarms in ICU as they may leverage both the machine learning model and nurse decision making. Ideally, according to the machine learning model output, nurses can implement decision-making via a user interface to tune the model in real-time. The collaboration between nurses and engineers to optimize personalized machine learning algorithms may provide an effective solution for this critical issue in the future. Intelligent algorithms can be developed to learn an individual’s baseline and provide timely feedback on the appropriate thresholds to guide personalized monitoring.

Conclusions

Clinical customizations of alarm settings on bedside physiological monitors helps to reduce excessive false alarms. Developed alarm algorithms to reduce false alarms in the ICU have achieved excellent performance in detecting and suppressing a single type of false alarm. Future collaboration between nurses and engineers to optimize personalized alarm algorithms has the greatest potential for false alarm reduction in the ICU.

Abbreviations

BP: 

backpropagation

CEASE: 

Communication, Electrodes, Appropriateness, and Setup alarm parameters, and Education

CINAHL: 

Cumulative Index to Nursing and Allied Health Literature

CNN: 

convolutional neural networks

DFO: 

dispersive files optimization

ECG: 

electrocardiogram

ERT: 

extremely randomized tree

ICU: 

intensive care units

PPG: 

photoplethysmogram

RF: 

random forest

SVM: 

support vector machine

Declarations

Acknowledgement

None.

Funding

None.

Conflict of interest

The authors have no conflicts of interest related to this publication.

Authors’ contributions

Study design: SFW, JR, AL; Performance of experiments: JYH, SFW, AL; Analysis, and interpretation of data: SFW, JYH, AL; Manuscript writing: JYH, SFW, JR, AL; Critical revision: JYH, SFW, JR, AL; Administration: SFW, AL. All authors have made a significant contribution to this study and have approved the final manuscript.

References

  1. Imhoff M, Kuhls S, Gather U, Fried R. Smart alarms from medical devices in the OR and ICU. Best Pract Res Clin Anaesthesiol 2009;23(1):39-50 View Article PubMed/NCBI
  2. Imhoff M, Kuhls S. Alarm algorithms in critical care monitoring. Anesth Analg 2006;102(5):1525-1537 View Article PubMed/NCBI
  3. Wung SF, Malone DC, Szalacha L. Sensory Overload and Technology in Critical Care. Crit Care Nurs Clin North Am 2018;30(2):179-190 View Article PubMed/NCBI
  4. Lehman EP, Krishnan RG, Zhao X, Mark RG, Lehman LH. Representation Learning Approaches to Detect False Arrhythmia Alarms from ECG Dynamics. Proc Mach Learn Res 2018;85:571-586 PubMed/NCBI
  5. Hever G, Cohen L, O’Connor MF, Matot I, Lerner B, Bitan Y. Machine learning applied to multi-sensor information to reduce false alarm rate in the ICU. J Clin Monit Comput 2020;34(2):339-352 View Article PubMed/NCBI
  6. Dai J, Sun Z, He X. Feature Engineering and Computational Intelligence in ECG Monitoring. Singapore: Springer; 2020 View Article
  7. Ghazanfari B, Zhang S, Afghah F, Payton-McCauslin N. Simultaneous multiple features tracking of beats: A representation learning approach to reduce false alarm rates in ICUs. Proceedings (IEEE Int Conf Bioinformatics Biomed) 2019;2019:2350-2355 View Article PubMed/NCBI
  8. Muroi C, Meier S, De Luca V, Mack DJ, Strässle C, Schwab P, et al. Automated False Alarm Reduction in a Real-Life Intensive Care Setting Using Motion Detection. Neurocrit Care 2020;32(2):419-426 View Article PubMed/NCBI
  9. Kasparick M, Golatowski F, Timmermann D. A safe and interoperable distributed alarm notification system for PoC medical devices using IEEE 11073 SDC. 2017 IEEE Healthcare Innovations and Point of Care Technologies (HI-POCT) 2017:257-261 View Article
  10. Atzema C, Schull MJ, Borgundvaag B, Slaughter GR, Lee CK. ALARMED: adverse events in low-risk patients with chest pain receiving continuous electrocardiographic monitoring in the emergency department. A pilot study. Am J Emerg Med 2006;24(1):62-67 View Article PubMed/NCBI
  11. Sendelbach S, Funk M. Alarm fatigue: a patient safety concern. AACN Adv Crit Care 2013;24(4):378-386 View Article PubMed/NCBI
  12. Fernandes C, Miles S, Lucena CJP. Detecting False Alarms by Analyzing Alarm-Context Information: Algorithm Development and Validation. JMIR Med Inform 2020;8(5):e15407 View Article PubMed/NCBI
  13. Wung SF. Bradyarrhythmias: Clinical Presentation, Diagnosis, and Management. Crit Care Nurs Clin North Am 2016;28(3):297-308 View Article PubMed/NCBI
  14. Ruppel H, Funk M, Kennedy HP, Bonafide CP, Wung SF, Whittemore R. Challenges of customizing electrocardiography alarms in intensive care units: A mixed methods study. Heart Lung 2018;47(5):502-508 View Article PubMed/NCBI
  15. Graham KC, Cvach M. Monitor alarm fatigue: standardizing use of physiological monitors and decreasing nuisance alarms. Am J Crit Care 2010;19(1):28-34 View Article PubMed/NCBI
  16. Cho OM, Kim H, Lee YW, Cho I. Clinical Alarms in Intensive Care Units: Perceived Obstacles of Alarm Management and Alarm Fatigue in Nurses. Healthc Inform Res 2016;22(1):46-53 View Article PubMed/NCBI
  17. Takla G, Petre JH, Doyle DJ, Horibe M, Gopakumaran B. The problem of artifacts in patient monitor data during surgery: a clinical and methodological review. Anesth Analg 2006;103(5):1196-1204 View Article PubMed/NCBI
  18. Schmid F, Goepfert MS, Reuter DA. Patient monitoring alarms in the ICU and in the operating room. Crit Care 2013;17(2):216 View Article PubMed/NCBI
  19. Hravnak M, Chen L, Dubrawski A, Bose E, Clermont G, Pinsky MR. Real alerts and artifact classification in archived multi-signal vital sign monitoring data: implications for mining big data. J Clin Monit Comput 2016;30(6):875-888 View Article PubMed/NCBI
  20. Hooman OM, Al-Rifaie MM, Nicolaou MA. Deep neuroevolution: training deep neural networks for false alarm detection in intensive care units. 2018 26th European Signal Processing Conference (EUSIPCO) 2018:1157-1161 View Article
  21. Ruppel H, Funk M, Whittemore R. Measurement of Physiological Monitor Alarm Accuracy and Clinical Relevance in Intensive Care Units. Am J Crit Care 2018;27(1):11-21 View Article PubMed/NCBI
  22. Lameski P, Zdravevski E, Koceski S, Kulakov A, Trajkovik V. Suppression of Intensive Care Unit False Alarms based on the Arterial Blood Pressure Signal. IEEE Access 2017;5:5829-5836 View Article
  23. Au-Yeung WM, Sahani AK, Isselbacher EM, Armoundas AA. Reduction of false alarms in the intensive care unit using an optimized machine learning based approach. NPJ Digit Med 2019;2:86 View Article PubMed/NCBI
  24. Li Q, Clifford GD. Signal quality and data fusion for false alarm reduction in the intensive care unit. J Electrocardiol 2012;45(6):596-603 View Article PubMed/NCBI
  25. Eerikäinen LM, Vanschoren J, Rooijakkers MJ, Vullings R, Aarts RM. Reduction of false arrhythmia alarms using signal selection and machine learning. Physiol Meas 2016;37(8):1204-1216 View Article PubMed/NCBI
  26. Xu J, Zhang Y, Zhang P, Mahmood A, Li Y, Khatoon S. Data Mining on ICU Mortality Prediction Using Early Temporal Data: A Survey. International Journal of Information Technology & Decision Making 2017;16:117-159 View Article
  27. Meyfroidt G, Güiza F, Ramon J, Bruynooghe M. Machine learning techniques to examine large patient databases. Best Pract Res Clin Anaesthesiol 2009;23(1):127-143 View Article PubMed/NCBI
  28. Lucas P. Bayesian analysis, pattern analysis, and data mining in health care. Curr Opin Crit Care 2004;10(5):399-403 View Article PubMed/NCBI
  29. Saeed M, Villarroel M, Reisner AT, Clifford G, Lehman LW, Moody G, et al. Multiparameter Intelligent Monitoring in Intensive Care II: a public-access intensive care unit database. Crit Care Med 2011;39(5):952-960 View Article PubMed/NCBI
  30. Stow PJ, Hart GK, Higlett T, George C, Herkes R, McWilliam D, et al. Development and implementation of a high-quality clinical database: the Australian and New Zealand Intensive Care Society Adult Patient Database. J Crit Care 2006;21(2):133-141 View Article PubMed/NCBI
  31. Peters MD, Godfrey CM, Khalil H, McInerney P, Parker D, Soares CB. Guidance for conducting systematic scoping reviews. Int J Evid Based Healthc 2015;13(3):141-146 View Article PubMed/NCBI
  32. Bi J, Yin X, Li H, Gao R, Zhang Q, Zhong T, et al. Effects of monitor alarm management training on nurses’ alarm fatigue: A randomised controlled trial. J Clin Nurs 2020;29(21-22):4203-4216 View Article PubMed/NCBI
  33. Ruppel H, Funk M, Whittemore R, Wung SF, Bonafide CP, Powell Kennedy H. Critical care nurses’ clinical reasoning about physiologic monitor alarm customization: An interpretive descriptive study. Journal of Clinical Nursing 2019;28:3033-3041 View Article
  34. Allan SH, Doyle PA, Sapirstein A, Cvach M. Data-Driven Implementation of Alarm Reduction Interventions in a Cardiovascular Surgical ICU. Jt Comm J Qual Patient Saf 2017;43(2):62-70 View Article PubMed/NCBI
  35. Wung SF, Schatz MR. Critical Care Nurses’ Cognitive Ergonomics Related to Medical Device Alarms. Crit Care Nurs Clin North Am 2018;30(2):191-202 View Article PubMed/NCBI
  36. Honan L, Funk M, Maynard M, Fahs D, Clark JT, David Y. Nurses’ Perspectives on Clinical Alarms. Am J Crit Care 2015;24(5):387-395 View Article PubMed/NCBI
  37. Fidler RL, Pelter MM, Drew BJ, Palacios JA, Bai Y, Stannard D, et al. Understanding heart rate alarm adjustment in the intensive care units through an analytical approach. PLoS One 2017;12(11):e0187855 View Article PubMed/NCBI
  38. Lewis CL, Oster CA. Research Outcomes of Implementing CEASE: An Innovative, Nurse-Driven, Evidence-Based, Patient-Customized Monitoring Bundle to Decrease Alarm Fatigue in the Intensive Care Unit/Step-down Unit. Dimens Crit Care Nurs 2019;38(3):160-173 View Article PubMed/NCBI
  39. Mäkivirta A, Koski E, Kari A, Sukuvaara T. The median filter as a preprocessor for a patient monitor limit alarm system in intensive care. Comput Methods Programs Biomed 1991;34(2-3):139-144 View Article PubMed/NCBI
  40. Zong W, Nielsen L, Gross B, Brea J, Frassica J. A practical algorithm to reduce false critical ECG alarms using arterial blood pressure and/or photoplethysmogram waveforms. Physiol Meas 2016;37(8):1355-1369 View Article PubMed/NCBI
  41. Li Q, Mark RG, Clifford GD. Robust heart rate estimation from multiple asynchronous noisy sources using signal quality indices and a Kalman filter. Physiol Meas 2008;29(1):15-32 View Article PubMed/NCBI
  42. Zong W, Moody GB, Mark RG. Reduction of false arterial blood pressure alarms using signal quality assessment and relationships between the electrocardiogram and arterial blood pressure. Med Biol Eng Comput 2004;42(5):698-706 View Article PubMed/NCBI
  43. Srivastava C, Sharma S, Jalali A. A novel algorithm for reducing false arrhythmia alarms in intensive care units. Annu Int Conf IEEE Eng Med Biol Soc 2016;2016:2525-2528 View Article PubMed/NCBI
  44. Krasteva V, Jekova I, Leber R, Schmid R, Abächerli R. Real-time arrhythmia detection with supplementary ECG quality and pulse wave monitoring for the reduction of false alarms in ICUs. Physiol Meas 2016;37(8):1273-1297 View Article PubMed/NCBI
  45. Pimentel MA, Santos MD, Springer DB, Clifford GD. Heart beat detection in multimodal physiological data using a hidden semi-Markov model and signal quality indices. Physiol Meas 2015;36(8):1717-1727 View Article PubMed/NCBI
  46. Antink CH, Leonhardt S. Reducing false arrhythmia alarms using robust interval estimation and machine learning. 2015 Computing in Cardiology Conference (CinC) 2015:285-288 View Article
  47. Silva P, Luz E, Silva G, Moreira G, Wanner E, Vidal F, et al. Towards better heartbeat segmentation with deep learning classification. Sci Rep 2020;10(1):20701 View Article PubMed/NCBI
  48. Mousavi S, Fotoohinasab A, Afghah F. Single-modal and multi-modal false arrhythmia alarm reduction using attention-based convolutional and recurrent neural networks. PLoS One 2020;15(1):e0226990 View Article PubMed/NCBI
  49. El Naqa I, Murphy MJ. Machine Learning in Radiation Oncology. Cham: Springer; 2015 View Article
  50. Clifford GD, Silva I, Moody B, Li Q, Kella D, Shahin A, et al. The PhysioNet/Computing in Cardiology Challenge 2015: Reducing False Arrhythmia Alarms in the ICU. Comput Cardiol (2010) 2015;2015:273-276 View Article PubMed/NCBI
  51. Hyland SL, Faltys M, Hüser M, Lyu X, Gumbsch T, Esteban C, et al. Early prediction of circulatory failure in the intensive care unit using machine learning. Nat Med 2020;26(3):364-373 View Article PubMed/NCBI
  52. Meyer J. Effects of warning validity and proximity on responses to warnings. Hum Factors 2001;43(4):563-572 View Article PubMed/NCBI
  53. Bitan Y, Meyer J, Shinar D, Zmora E. Nurses’ reactions to alarms in a neonatal intensive care unit. Cogn Tech Work 2004;6(4):239-246 View Article
  54. Bitan Y, O’Connor MF. Correlating data from different sensors to increase the positive predictive value of alarms: an empiric assessment. F1000Res 2012;1:45 View Article PubMed/NCBI
  • Exploratory Research and Hypothesis in Medicine
  • pISSN 2993-5113
  • eISSN 2472-0712
Back to Top

Reducing False Alarms in Intensive Care Units: A Scoping Review

Jiayan Huo, Shufen Wung, Janet Roveda, Ao Li
  • Reset Zoom
  • Download TIFF